TERUG NAAR GL820
SPECIFICATIES
|
|

GL820 SPECIFICATIES
| item |
Description |
| Number of analog input
channels |
20 ch, Expandable up to 200 ch by
unit of 20 ch |
| External input output |
Input *1 |
Trigger or Sampling input 1ch,
Logic or Pulse input 4 ch |
| Output *1 |
Alarm output 4 ch |
| Sampling interval |
10 ms to 1 h (in 10 ms to 50 ms,
voltage only and limited channel), External |
| Time scale |
1 sec to 24 hour/division |
| Trigger function |
Action |
Start or stop capturing data by the
trigger |
| Source |
Start: Off, Input signal, Alarm,
External *1, Clock, Week or Time |
| Stop: Off, Input signal, Alarm,
External *1, Clock, Week or Time |
| Combination |
OR or AND condition at the level of
signal or edge of signal |
| Condition |
Analog: Rising, Falling, Window-in,
Window-out |
| Pulse: Rising, Falling, Window-in,
Window-out |
| Logic: Rising, Falling |
| Alarm function |
Detecting method |
Level or edge of signal |
| Condition |
Analog: Rising, Falling, Window-in,
Window-out |
| Pulse: Rising, Falling, Window-in,
Window-out |
| Logic: Rising, Falling |
| Alarm output *1 |
4 channels, Output type: Open
collector (pulled-up to 5 V by resistor 10 kΩ) |
| Pulse input function
*1 |
Accumulating count mode |
Accumulating the number of pulses
from the start of measurement
Range: 50, 500, 5k, 50k, 500k, 5M, 50M, 500M counts/F.S. |
| Instant count mode |
Counting the number of pulses per
sampling interval
Range: 50, 500, 5k, 50k, 500k, 5M, 50M, 500M counts/F.S. |
| Rotation count (RPM) mode |
Counting the number of pulses per
second and then it is converted to RPM
Range: 50, 500, 5k, 50k, 500k, 5M, 50M, 500M rpm /F.S. |
| Max. input pulse rate |
50k pulses/sec or 50k counts per
sampling interval (16 bits counter is used) |
| Calculation function |
Between channels |
Addition, Subtraction,
Multiplication and Division for analog input |
| Statistical |
Select two calculations from
Average, Peak, Max., Min., RMS |
| Search function |
Search for analog signal levels,
values of logic or pulse or alarm point in captured data |
| Interface to PC |
Ethernet (10 BASE-T/100 BASE-TX),
USB (Full speed) |
| Storage device |
Built-in Flash memory (2 giga-bytes),
USB memory device *2 |
| Data saving function |
Captured data |
Direct saving of data into built-in
Flash memory or USB memory device |
| Others |
Setting conditions, Screen copy |
| Ring capturing mode |
Function: ON/OFF, Number of
capturing point: 1000 to 2000000 (size of the capture data will be
limited to 1/3 of available memory when in Ring Mode) |
| USB memory device
emulation |
USB Memory emulation mode (Transfer
or delete the file in built-in memory) |
| Engineering scale
function |
Set based on the reference point of
the scaled output and input signal for each channel (Voltage
measurement: four points are necessary to scale the output,
Temperature measurement: two points are necessary to scale the
output). |
| Display |
Size |
5.7 inch TFT color LCD (VGA: 640 x
480 dots) |
| Formats |
Waveform + Digital, Waveform only,
Calculation + Digital, Expanded digital |
| Operating environment |
0 °C to 45°C, 5 % to 85 % RH
(When operating with battery pack 0 °C to 40 °C, charging battery 15
°C to 35°C) |
| Power source |
AC adapter (100 V to 240 V, 50/60
Hz)
DC power (8.5 V to 24 V DC, max. 26.4 V)*3, Battery pack*3 |
| Power consumption |
32VA or lower
(When operating with AC adapter, displaying LCD, charging battery
pack) |
| External dimensions (WxDxH) |
approx. 232 x 152 x 50 mm |
| Weight |
approx. 900 g (Excluding AC adapter
and battery pack) |
Software specifications
| Item |
Description |
| Supported OS |
Windows XP / Vista / 7 (32 bits and
64 bits edition) |
| Functions |
Control GL820, Real-time data
capture, Replay data, Data format conversion |
| GL820 setting control |
Input settings, Memory settings,
Alarm settings, Trigger settings |
| Controlled units |
Up to 10 units or 500 channels |
| Captured data |
Transfers data in real-time (in
binary or CSV format), saved data in GL820 or the USB memory |
| Displayed information |
Analog waveforms, Logic waveforms,
Pulse waveforms, Digital values |
| Display modes |
Y-T waveforms, Digital values,
Report, X-Y graph (specified period of data, data reply only) |
| Warning functions |
Sends E-mail to the specified
address when the alarms occur |
| File format conversions |
Converts the specified period data
or all data to the CSV format (thinning function is available) |
| Report functions |
Creates the daily or monthly report
automatically (can also export directly to Excel) |
Standard accessories
| Item |
Description |
Quantity |
| AC adapter |
100 V to 240 V AC, 50/60 Hz (with
specified type of power cord) |
1set |
| CD-ROM |
User's manual (PDF format),
Application software |
1piece |
| Quick Start Guide |
|
1copy |
Analog input specifications
| Item |
Description |
| Type of input terminal |
Screw terminal (M3
screw) |
| input method |
Scans by the
photo-MOS-relay, all channels isolated, balanced input |
| Measurement range |
Voltage
Temperature |
20, 50, 100, 200, 500
mV, 1, 2, 5, 10, 20, 50 V, and 1-5 V /F.S.
Thermocouple: K, J, E, T, R, S, N, and W (WRe5-26) |
| Resistance Temperature
Detectors (RTDs): Pt100, JPt100 (JIS), Pt1000 (IEC751) |
| Humidity |
0 to 100%
(using humidity sensor (B-530 optional), power is supplied to only
one sensor) |
| Filter |
Off, 2, 5, 10, 20, 40 (moving
average in selected number) |
| Measurement accuracy
*4 |
Voltage |
0.1% of F.S. |
|
Temperature |
Thermocouple |
Measurement range |
Measurement accuracy |
|
R/S |
0 °C ≦ TS ≦ 100 °C
100 °C ≦ TS ≦ 300 °C
R: 300 °C ≦ TS ≦ 1600 °C
S: 300 °C ≦ TS ≦ 1760 °C |
±5.2 °C
±3.0 °C
±(0.05 % of reading +2.0 °C)
±(0.05 % of reading +2.0 °C) |
|
B |
400 °C ≦ TS ≦ 600 °C
600 °C ≦ TS ≦ 1820 °C |
±3.5 °C
±(0.05% of reading +2.0 °C) |
|
K |
-200 °C ≦ TS ≦ -100 °C
-100 °C ≦ TS ≦ 1370 °C |
±(0.05 % of reading +2.0 °C)
±(0.05 % of reading +1.0 °C) |
|
E |
-200 °C ≦ TS ≦ -100 °C
-100 °C ≦ TS ≦ 800 °C |
±(0.05 % of reading +2.0 °C)
±(0.05 % of reading +1.0 °C) |
|
T |
-200 °C ≦ TS ≦ -100 °C
-100 °C ≦ TS ≦ 400 °C |
±(0.1 % of reading +1.5 °C)
±(0.1 % of reading +0.5 °C) |
|
J |
-200 °C ≦ TS ≦ -100 °C
-100 °C ≦ TS ≦ 100 °C
100 °C ≦ TS ≦ 1100 °C |
±2.7 °C
±1.7 °C
±(0.05 % of reading +1.0 °C) |
|
N |
0 °C ≦ TS ≦ 1300 °C |
±(0.1 % of reading +1.0 °C) |
|
W |
0 °C ≦ TS ≦ 2000 °C |
±(0.1 % of reading +1.5 °C) |
| |
Reference Junction
Compensation (R.J.C.): ±0.5 °C |
|
RTD |
Measurement range |
Measurement accuracy |
|
Pt100 |
-200 °C to 850 °C (FS=1050 °C) |
±1.0 °C |
|
JPt100 |
-200 °C to 500 °C (FS=700 °C) |
±0.8 °C |
|
Pt1000 |
-200 °C to 500 °C (FS=700 °C) |
±0.8 °C |
| A/D Converter |
ΣΔtype, 16 bits (effective
resolution: 1/40000 of measuring full range) |
| Maximum input voltage |
Between +/- terminal |
60 Vp-p |
| Between channels |
60 Vp-p |
| Between channels / GND |
60 Vp-p |
| Withstand voltage |
Between channels |
350 Vp-p (1minute) |
| Between channels(-) /
GND |
350 Vp-p (1minute) |
- *1
- Logic alarm cable (B-513) option is
required.
Input signal of External sampling, Logic, Pulse; Maximum voltage: 24 V,
Threshold: approx. 2.5 V, Hysteresis: approx. 0.5 V
- *2
- Size of the USB memory device is
unlimited. Maximum file size is limited to 2 GB
- *3
- DC drive cable (B-514) or battery pack
(B-517) option is required.
- *4
- Subject to the following conditions;
- Room Temperature is 23 °C±5 °C
- When 30 minutes or more have
elapsed after power was turned on.
- Filter is set to 10
- Sampling rate is set to 1s with 20
channels.
- GND terminal is connected to ground.
|