GL220
SPECIFICATIES
|
|

GL220 SPECIFICATIES
|
Item |
Description |
|
Number of analog input
channels |
10 ch |
|
External input/output |
Input *1 |
Trigger or Sampling input 1
ch, Logic or Pulse input 4 ch |
|
Output *1 |
Alarm output 4 ch |
|
Sampling interval |
10 ms to 1 h (in 10ms to
50ms, voltage only and limited channel), External |
|
Time scale |
1 sec to 24 hour /division |
|
Trigger function |
Action |
Start or stop capturing data
by the trigger |
|
Source |
Start: Off, Input signal,
Alarm, External *1 , Clock, Week or Time |
|
Stop: Off, Input signal,
Alarm, External *1 , Clock, Week or Time |
|
Combination |
OR or AND condition at the
level of signal or edge of signal |
|
Condition |
Analog: Rising, Falling,
Window-in, Window-out |
|
Pulse: Rising, Falling,
Window-in, Window-out |
|
Logic: Rising or Falling |
|
Alarm function |
Detecting method |
Level or edge of signal |
|
Condition |
Analog: Rising, Falling,
Window-in, Window-out |
|
Pulse: Rising, Falling,
Window-in, Window-out |
|
Logic: Rising, Falling |
|
Alarm output *1 |
4 channels, Output type: Open
collector (pull-up resistor 10 kΩ) |
|
Pulse input function *1 |
Accumulating count mode |
Accumulating the number of
pulses from the start of measurement
Range: 50, 500, 5 k, 50 k, 500 k, 5 M, 50 M, 500 M counts/F.S. |
|
Instant count mode |
Counting the number of pulses
per sampling interval
Range: 50, 500, 5 k, 50 k, 500 k, 5 M, 50 M, 500 M counts/F.S. |
|
Rotation count (RPM) mode |
Counting the number of pulses
per second and then it is converted to RPM
Range: 50 rpm, 500 rpm, 5 krpm, 50 krpm, 500 krpm, 5 Mrpm, 50 Mrpm,
500 Mrpm /F.S. |
|
Max. input pulse rate |
50 k pulses/sec or 50k counts
per sampling interval, 16 bits counter is used |
|
Calculation function |
Between channels |
Addition, Subtraction,
Multiplication and Division for analog input |
|
Statistical |
Select two calculations from
Average, Peak, Max., Min., RMS |
|
Search function |
Search for analog signal
levels, values of logic or pulse or alarm point in captured data |
|
Interface to PC |
USB (Full speed) |
|
Storage device |
Built-in Flash memory (2
giga-bytes), USB memory device *2 |
|
Data saving function |
Captured data |
Direct saving of data into
built-in Flash memory or USB memory device |
|
Others |
Setting conditions, Screen
copy |
|
Ring capturing mode |
Function: ON/OFF, Number of
capturing point: 1,000 to 2,000,000
(Size of the capture data will be limited to 1/3 of available memory) |
|
USB memory device emulation |
USB Memory emulation mode
(Transfer or delete the file in built-in memory) |
|
Engineering scale function |
Set based on the reference
point of the scaled output and input signal for each channel
(Voltage measurement: four points are necessary to scale the output,
Temperature measurement: two points are necessary to scale the
output). |
|
Display |
Size |
4.3 inch TFT color LCD (WQVGA:
480 x 272 dots) |
|
Formats |
Waveform + Digital, Waveform
only, Calculation + Digital, Expanded digital |
|
Operating environment |
0 to 45 °C, 5 to 85 %RH
(When operating with battery pack 0 to 40 °C, charging battery 15 to
35 °C) |
|
Power source |
AC adapter (100 to 240 V,
50/60 Hz), DC power (8.5 to 24 V DC, max. 26.4 V) *3 ,
Battery pack *3 |
|
Power consumption |
29 VA or lower (when
operating with AC adapter, displaying LCD) |
|
External dimensions (W×D×H) |
approx. 194 x 117 x 42 mm |
|
Weight |
approx. 520 g (Excluding AC
adapter and battery pack) |
- *1
- Logic alarm cable (B-513)
option is required.
Input signal of External sampling, Logic, Pulse; Maximum voltage: 24 V,
Threshold: approx. 2.5 V, Hysteresis: approx. 0.5 V.
- *2
- Size of the USB memory device
is unlimited. Maximum file size is limited to 2GB.
- *3
- DC drive cable (B-514) or
battery pack (B-517) option is required.
Analog input specifications
| Item |
Description |
| Type of input
terminal |
Screw
terminal (M3 screw) |
| Input method |
Scans by the
photo-MOS-relay, all channels isolated, balanced input |
| Measurement
range |
Voltage |
20, 50, 100,
200, 500 mV, 1, 2, 5, 10, 20, 50 V, and 1-5 V /F.S. |
| Temperature |
Thermocouple:
K, J, E, T, R, S, B, N, and W (WRe5-26) |
| Humidity |
0 to 100% (using
humidity sensor (B-530 optional), power is supplied to only one
sensor) |
| Filter |
Off, 2, 5,
10, 20, 40 (moving average in selected number) |
| Measurement
accuracy *4 |
Voltage |
0.1 % of F.S. |
|
Temperature |
Thermocouple |
Measurement range |
Measurement Accuracy |
|
R/S |
0 °C ≤ TS ≤ 100 °C
100 °C < TS ≤ 300 °C
R: 300 °C < TS ≤ 1600 °C
S: 300 °C < TS ≤ 1760 °C |
± 5.2 °C
± 3.0 °C
± (0.05 % of reading + 2.0 °C)
± (0.05 % of reading + 2.0 °C) |
|
B |
400 °C ≤ TS ≤ 600 °C
600 °C < TS ≤ 1820 °C |
± 3.5 °C
± (0.05 % of reading + 2.0 °C) |
|
K |
-200 °C ≤ TS ≤ -100 °C
-100 °C < TS ≤ 1370 °C |
± (0.05 % of reading +
2.0 °C)
± (0.05 % of reading + 1.0 °C) |
|
E |
-200 °C ≤ TS ≤ -100 °C
-100 °C < TS ≤ 800 °C |
± (0.05 % of reading +
2.0 °C)
± (0.05 % of reading + 1.0 °C) |
|
T |
-200 °C ≤ TS ≤ -100 °C
-100 °C < TS ≤ 400 °C |
± (0.1 % of reading + 1.5
°C)
± (0.1 % of reading + 0.5 °C) |
|
J |
-200 °C ≤ TS ≤ -100 °C
-100 °C < TS ≤ 100 °C
100 °C < TS ≤ 1100 °C |
± 2.7 °C
± 1.7 °C
± (0.05 % of reading + 1.0 °C) |
|
N |
0 °C ≤ TS ≤ 1300 °C |
± (0.1 % of reading + 1.0
°C) |
|
W |
0 °C ≤ TS ≤ 2000 °C |
± (0.1 % of reading + 1.5
°C) |
| |
Reference
Junction Compensation (R.J.C.): ±0.5 °C |
| A/D converter |
ΣΔ type, 16
bits (effective resolution: 1/40,000 of measuring full range) |
| Maximum input
voltage |
Between + / -
terminal |
60 V p-p |
| Between
channels |
60 V p-p |
| Between
channel / GND |
60 V p-p |
| Withstand
voltage |
Between
channels |
350 V p-p (1
minute) |
| Between
channel(-)/ GND |
350 V p-p (1
minute) |
- *4
- Subject to the following
conditions;
• Room Temperature is 23°C ±5°C.
• When 30 minute or more have elapsed after power was turned on.
• Filter is set to 10.
• Sampling rate is set to 1s with 10 channels.
• GND terminal is connected to the ground.
Software specifications
| Item |
Description |
| Supported OS |
Windows XP / Vista / 7
(32 bits and 64 bits edition) |
| Functions |
Control GL220, Real-time
data capture, Replay data, Data format conversion |
| GL220 settings control |
Input settings, Memory
settings, Alarm settings, Trigger settings |
| Captured data |
Transfers data in
real-time (in binary or CSV format), saved data in GL220 or the USB
memory |
| Displayed information |
Analog waveforms, Logic
waveforms, Pulse waveforms, Digital values |
| Display modes |
Y-T waveforms, Digital
values, Report, X-Y graph (specified period of data, data replay
only) |
| Warning functions |
Sends E-mail to the
specified address when the alarm occurred |
| File format conversions |
Converts the specified
period data or all data to the CSV format (thinning function is
available) |
| Report functions |
Creates a daily or
monthly report automatically (can also export directly to Excel) |
| Displayed Max. Min. |
Displays the maximum,
minimum and current value in measurement |
Standard accessories
| Item |
Description |
Quantity |
| AC adapter |
100 to 240 V AC, 50 / 60
Hz (with specified type of power cord) |
1 set |
| CD-ROM |
User’s manual (PDF format),
Application software |
1 piece |
| Quick Start Guide |
|
1 copy |
|